Knowledge base  /  Documentation
Understanding documentation

How do you read a chromatogram?

From retention time to peak area: read a chromatogram alongside the method and the original analytical report.
Novogin knowledge base · Guidance for laboratory research

A chromatogram shows how the measured detector signal changes during a separation. The horizontal axis usually shows time; the vertical axis shows the signal from the detector used. Peaks indicate when something produced a measurable signal. Without the method, detector, sample information and peak assignment, the graph alone does not tell you which substances are present or how much of them is in the entire container.

Start with the axes and the analytical method

In liquid chromatography, the separated sample leaves the column over time. The detector records a signal, which is plotted as a line. Waters describes a chromatogram as a representation of this separation: peaks rise above a baseline as components pass the detector.

First check the caption and method details. Does the analysis use HPLC with UV detection, a mass detector or another configuration? With UV detection, the selected wavelength affects what is visible. Chromatograms from different detectors or settings are not directly comparable, even when the same sample was used. HPLC and LC-MS are related, but their detectors answer different analytical questions.

Next, read the units on both axes. A time axis in minutes and a vertical axis in absorbance units require a different interpretation from an ion signal from a mass spectrometer. The scale may have been set automatically. A tall peak on the graph is therefore not a standalone measure of the quantity in a vial.

What does a peak tell you?

The retention time is the time at which a peak emerges from the column under the conditions used. Its position can help with comparison against a reference substance, but the same retention time alone does not identify a substance with certainty. Other substances may elute nearby. An assignment needs a documented method and, where necessary, additional identity information.

The peak height is the highest signal above the baseline. The peak area is the integrated area under the peak. Waters explains that quantification generally uses peak area. The boundaries the software uses for the start and end of a peak affect the calculation. Check whether the report includes an integration table, peak names and a method description. An image without this context provides limited evidence.

A flat section between peaks is called the baseline. An unstable baseline, peak shoulders or strongly overlapping peaks make interpretation more difficult. The separation between two peaks is often described in terms of resolution. How clearly two components are separated matters before their areas are compared.

Why one large peak does not prove everything

A chromatogram with one dominant peak may look convincing, but two components can leave the column at the same time. This is called co-elution. Waters describes incompletely separated components that appear as one unresolved peak. An Agilent application note shows that mass detection can reveal hidden components beneath a UV peak. That application concerns a different analyte; the principle is that a UV trace alone cannot rule out every instance of co-elution.

The method must therefore be suitable for the question. ICH Q2(R2) addresses selectivity and describes how complementary procedures based on different principles can be useful when a single procedure does not provide sufficient discrimination. This is general analytical methodology, not a claim that a particular Novogin report used two techniques.

What does area% mean in the peak table?

Some reports express the area of a peak as a percentage of the sum of the included peak areas. Agilent calls this area%. The calculation requires no prior calibration, but compares signals within the selected analysis. Without correcting for differences in detector response, area% is not an accurate mass percentage of all substances in the sample.

An example unrelated to any actual Novogin report: if the main peak accounts for 99 out of 100 integrated area units, the software may show 99 area%. This describes what was counted within that detector trace and those integration rules. It does not prove that the entire vial contains 99% target substance by mass, or that the nominal milligram quantity was measured. Identity, purity and measured content must be interpreted separately.

A practical reading sequence

  1. Link the document to the sample. Read the product, batch, sample ID and report number. Use the COA guide if these fields are unfamiliar. Then check whether the report matches the current batch.
  2. Identify the type of chromatogram. Note the method, detector, wavelength or measurement mode, axes and units.
  3. Find the peak assignment. Which peak does the laboratory identify as the target substance, and what comparison supports that assignment?
  4. Check the separation. Are neighbouring peaks clearly distinguishable? Is there a shoulder or possible overlap?
  5. Read the calculation. Does it report area, area%, a calibrated quantity or another result? Which peaks were included?
  6. Read the conclusion in the context of the full report. Only accept properties for which the procedure, result and sample linkage together provide sufficient information.

A chromatogram is an important part of a certificate of analysis, but it is not a standalone certificate. The COA library provides access to available original reports. Ask for clarification if peak assignments are unclear or measurement details are missing, rather than treating the graph as evidence for more than it can demonstrate.

Sources

  1. Waters, Bands, Peaks and Band Spreading — baseline, time axis and peaks.
  2. Waters, Appendix: HPLC Nomenclature — retention time, area and resolution.
  3. Agilent OpenLab, Area% and height% — calculation and limitations of area%.
  4. Agilent, Mass spectrometry detection enables identification of coeluting compounds — an example of co-elution in UV and MS detection.
  5. EMA / ICH Q2(R2), Validation of analytical procedures — selectivity and fitness for the analytical purpose.

This knowledge base covers analytical documentation for professional laboratory research. It does not provide advice on use in humans or animals.

From guidance to documentation

View the details of your batch.

Find the report that matches your product, strength and batch number.