Certificates of analysis often mention HPLC and LC-MS. These techniques provide different types of information about the same sample. This guide mainly compares HPLC with UV detection and LC-MS. LC-MS also uses liquid chromatography: the distinction partly lies in the detector and the question the laboratory is investigating.
HPLC with UV detection compared with LC-MS
| Question | HPLC with UV detection | LC-MS |
|---|---|---|
| What is measured? | The UV signal of separated components over time. | Ionic m/z values after chromatographic separation. |
| What can this help assess? | The peak pattern and, within the method used, relative peak areas. | Whether the measured m/z values match the expected mass. |
| What does this not establish on its own? | A single peak or high percentage alone does not establish identity or the amount in milligrams. | A matching mass alone cannot confirm the complete amino acid order. |
A quantitative result requires a suitable measurement setup with either technique. Read the reported result, unit and calibration; do not infer an amount from a graph or method name alone.
What does HPLC do?
HPLC stands for High Performance Liquid Chromatography. A liquid carries the sample through a column. Compounds travel through it differently and can therefore be separated. A detector measures what leaves the column; the chromatogram plots the measured signal against time.
Source: Waters — how HPLC and detectors work
With UV detection, the signal depends on the wavelength selected. Also consider how well the peaks are separated. Compounds leaving the column at almost the same time can produce overlapping peaks. A single prominent peak is therefore not, by itself, a complete identification of a peptide.
For more detail, see the Dutch article on reading a chromatogram, including axes, peaks and the limits of interpretation.
What does LC-MS add?
LC-MS couples the separation to mass spectrometry. The mass spectrometer measures the mass-to-charge ratio of ions, written as m/z. A peptide can carry multiple charges, so different m/z signals may belong to the same peptide. The laboratory takes those charges into account when comparing the results with the expected molecular mass.
Source: Waters — peptide analysis with UV and mass detection
A matching mass supports identification, but it is not the same as establishing the complete amino acid sequence. Fragmentation analysis, such as MS/MS, can provide additional structural information. Ask which identity checks were actually performed; the label LC-MS alone does not give the full answer.
Source: Waters — peptide mapping and structural information
What does this tell you about purity and quantity?
An HPLC purity percentage may be based on the area of the assigned peptide peak relative to the peaks included in the calculation. Check which calculation and detection method were used. That percentage does not automatically describe the entire contents of the vial: water and salts, for example, are not accounted for in the same way in a UV peptide analysis.
A report showing 99% chromatographic purity therefore does not, by itself, prove that a vial contains the number of milligrams stated on its label. Purity and measured quantity are separate measurements.
Source: Sigma-Aldrich — peptide purity and peptide content
Both HPLC and LC-MS can form part of a quantitative assay. This requires an appropriate measurement approach, such as calibration with standards containing known amounts. Look in the report for the quantitative result, its unit and the basis of the calculation. A graph or method name does not replace those details.
Source: Agilent — quantitation and calibration
In identity, purity and measured quantity (in Dutch) you can read why these results answer separate questions.
Five checks for your report
- Check that the product, variant and batch number match your documentation.
- Check which technique and detector were used; do not assume additional tests were performed.
- Keep reported identity, purity and quantity separate.
- Check the units, test date and any explanation of discrepancies.
- Ask the laboratory or supplier for clarification if a result is unclear.
This guide helps you read a report; it is not an assessment of a specific Novogin batch. Do not infer a sterility result or suitability for human use from an HPLC or LC-MS label. Draw conclusions only about properties that were actually tested and reported.
Further reading and reports
Read also: how do you read a COA?
See also the Dutch guide on checking whether an analysis report matches the current batch.
Browse available reports in the batch library
Would you like to compare product specifications alongside this explanation? Explore research peptides for professional research.
This knowledge base explains analytical documentation for laboratory research. It does not provide advice on dosing, administration or medical use.
Sources
- Waters — How Does High Performance Liquid Chromatography Work?: separation, detectors and chromatograms.
- Waters — Mass-Directed Isolation of Synthetic Peptides Using the Waters SQ Detector 2 (2024): mass-to-charge ratios and multiple charge states.
- Waters — Peptide Mapping and Analysis: structural information in a peptide-mapping workflow; this is not necessarily the same test as every LC-MS report.
- Sigma-Aldrich — Peptide Sample Amount Determination: peptide purity and peptide content.
- Agilent OpenLab CDS 2.8 — Glossary: see Calibration and Calibration standard for the relationship between instrument response and a known amount.